Analysis of Heart Pulse Transmission Parameters Determined from Multi-Channel PPG Signals Acquired by a Wearable Optical Sensor

Authors

  • Jiří Přibil Institute of Measurement Science, Slovak Academy of Sciences, Dúbravská cesta 9, 841 04 Bratislava, Slovakia https://orcid.org/0000-0002-7488-1643
  • Anna Přibilová Institute of Measurement Science, Slovak Academy of Sciences, Dúbravská cesta 9, 841 04 Bratislava, Slovakia https://orcid.org/0000-0002-5107-2552
  • Ivan Frollo Institute of Measurement Science, Slovak Academy of Sciences, Dúbravská cesta 9, 841 04 Bratislava, Slovakia https://orcid.org/0000-0002-8976-3308

DOI:

https://doi.org/10.2478/msr-2023-0028

Keywords:

arterial blood pressure, estimation by linear regression, photoplethysmography signal, pulse transmission time, wearable sensor

Abstract

The article describes the development and testing of a special prototype wearable device consisting of three optical photoplethysmography (PPG) sensors. The functionality of the developed triple PPG sensor was tested under normal laboratory conditions and in a running magnetic resonance imaging (MRI) scanner working with a low magnetic field. The results of the first measurements under normal laboratory conditions show that the obtained mutual positions of systolic/diastolic blood pressure values and heart pulse transmission parameters determined from the PPG waves can be fitted by a line segment with a sufficiently high slope. Measurement experiments inside the open-air MRI tomograph show the practical influence of vibrations and acoustic noise on the cardiac system of the examined persons, which was confirmed by a slight increase in the heart pulse rate and changes in pulse transmission time and pulse wave velocity. We plan to perform further measurements inside the whole-body MRI device producing more intensive vibrations and noise with expected higher stress impact on an exposed person.

Downloads

Published

17.10.2023

How to Cite

Analysis of Heart Pulse Transmission Parameters Determined from Multi-Channel PPG Signals Acquired by a Wearable Optical Sensor. (2023). Measurement Science Review, 23(5), 217-226. https://doi.org/10.2478/msr-2023-0028

Similar Articles

21-30 of 90

You may also start an advanced similarity search for this article.