Analysis of Heart Pulse Transmission Parameters Determined from Multi-Channel PPG Signals Acquired by a Wearable Optical Sensor

Authors

  • Jiří Přibil Institute of Measurement Science, Slovak Academy of Sciences, Dúbravská cesta 9, 841 04 Bratislava, Slovakia https://orcid.org/0000-0002-7488-1643
  • Anna Přibilová Institute of Measurement Science, Slovak Academy of Sciences, Dúbravská cesta 9, 841 04 Bratislava, Slovakia https://orcid.org/0000-0002-5107-2552
  • Ivan Frollo Institute of Measurement Science, Slovak Academy of Sciences, Dúbravská cesta 9, 841 04 Bratislava, Slovakia https://orcid.org/0000-0002-8976-3308

DOI:

https://doi.org/10.2478/msr-2023-0028

Keywords:

arterial blood pressure, estimation by linear regression, photoplethysmography signal, pulse transmission time, wearable sensor

Abstract

The article describes the development and testing of a special prototype wearable device consisting of three optical photoplethysmography (PPG) sensors. The functionality of the developed triple PPG sensor was tested under normal laboratory conditions and in a running magnetic resonance imaging (MRI) scanner working with a low magnetic field. The results of the first measurements under normal laboratory conditions show that the obtained mutual positions of systolic/diastolic blood pressure values and heart pulse transmission parameters determined from the PPG waves can be fitted by a line segment with a sufficiently high slope. Measurement experiments inside the open-air MRI tomograph show the practical influence of vibrations and acoustic noise on the cardiac system of the examined persons, which was confirmed by a slight increase in the heart pulse rate and changes in pulse transmission time and pulse wave velocity. We plan to perform further measurements inside the whole-body MRI device producing more intensive vibrations and noise with expected higher stress impact on an exposed person.

Downloads

Published

17.10.2023

How to Cite

Přibil, J., Přibilová, A., & Frollo, I. (2023). Analysis of Heart Pulse Transmission Parameters Determined from Multi-Channel PPG Signals Acquired by a Wearable Optical Sensor. Measurement Science Review, 23(5), 217–226. https://doi.org/10.2478/msr-2023-0028